Setting up the Hardware

beam-pipe. 2

Ion chamber 3

PF2S2. 4

Mar table. 4

Cryojet 5

Amptek fluorescence detector 7

Cameras. 8

 

beam-pipe

To collect macromolecular crystallography data in 5ID-B the beam transport pipe must first be in place.  The beam pipe allows us to minimize the air in the path of the X-ray beam.  Two pipes are used for the mar165: a short extension metal pipe (labeled “… mar165 1”) + a longer PVC pipe (labeled “… mar165 2”). 

If you are switching over from users in 5ID-C or 5ID-D, follow the instructions in mar165_protocol.  Do not do this switch unless you have been trained to do so.


Ion chamber

An ion chamber is installed in 5ID-B downstream of the slit tank, near the front of the hutch (see PF2S2 for picture).  Check the ion chamber connections as per this figure:


PF2S2

The PS2S2 / PF4 assembly should be immediately downstream of the ion chamber as shown in this figure.

In protoscan this assembly is referred to as the PF2S2 and contains an X-ray shutter and four filters.  Currently only 4 of the filters, containing Al plates of varied thickness, are used

Mar table

The Mar table is the bread-board table on wheels at the back of the 5ID-B hutch.  This table contains the Mar165 base, the Mar165 CCD detector, the Cryojet, and the Amptek detector.  The table itself is aligned with the beam.  The equipment is mounted on sliders which allow it to be translated out of the beam path without losing the alignment as per the  mar165_protocol.


Cryojet

The Cryojet is the Oxford system for keeping crystal samples at low temperatures during data collection.  It is mounted on a slider with the Amptek detector on the Mar table.  The controller for the Cryojet is under the Mar table.  When not in use, the rates for the cold “sample flow” (inner tube) and the warm “shield flow” (outer tube)  are turned down to 2.5 and 2.5 as shown in this figure.

this causes the sample temperature to rise above its set point.  During use, the flows are adjusted to 7.0 – 7.5 for the sample flow and 5.5 – 6.0 for the shield flow.  The flow and temperature are set by pressing and holding the respective black “set” button and adjusting the level with the red “raise” and “lower” buttons.


The Cryojet should be aligned collinear with the spindle axis.  This is best done from the backside of the mar table with a second person checking the alignment from the sample side.


Amptek fluorescence detector

If you will be collecting MAD data, you will need the Amptek detector to select your edge energy.  A step-by-step wiring checklist for this detector is available on the beam-line.  You can confirm the wiring by comparing it to the diagram below (also available on the beam-line).


Cameras

There are several video cameras in the 5ID-B hutch.  These are connected with BNC cables through the patch panel to the ProVideo Auto switch on the outside of the hutch.  The current camera configuration is as follows (referring to the channel numbers on the ProVideo switch):

1.      crystal
      located in the mar base underneath the crystal

2.      Cryojet nozzle
      on the Cryojet slider stage

3.      Cryojet controller
      located under the mar table

4.      Phi axis
      located on the rod to the right of the phi axis knobs

5.      variable
      usually located in the front of 5ID-B

6.      overview
      located on the top of the patch panel

7.      empty

8.      empty

Additionally the output from the ProVideo is also directed to a video card in banana1 (the SGI O2).  This allows you to use the program mediarecorder to capture images of whatever is on screen.